flash and fire point tester of petroleum products (Cleveland open cup method)

flash and fire point tester of petroleum products (Cleveland open cup method)
 

Product description:


FDT-0101 opening flash point and ignition tester
Applied standards:
GB/ T3536, ISO2592, ASTM D92
Application range
The instrument is designed and manufactured in accordance with the requirements stipulated by the People's Republic of standard GB/ T3536 petroleum products Determination of flash and fire points - Cleveland open cup method', applicable to the determination in addition to fuel oil and opening a flash point of less than 79 ° C. flash and fire points of petroleum products and asphalt.
Function Features
1, Using a special form of the heating furnace, to ensure the security of the test; heating power is continuously adjustable to meet the requirements of the test.
2, instrument is designed for desktop structure, stainless steel countertops and stainless steel furnace shell, nice.
3, instrument is complete, users only accompanied by gas or other civil combustible gas can be tested.
4, with the the flame extinguished cover, ignition test requirements.
Technical Specification
1 Power supply: AC 220V ± 10%, 50Hz.
2. heating device: quartz tube furnace heating, no open flame, explosion-proof, power 0 ~ 600W continuously adjustable.
3 zoned sweeping device: automatic scanning program.
4, thermometer: -6 to 400 ℃ indexing 2 ° C, the technical conditions are met petroleum products testing the liquid thermometer technical conditions GB/ T514.
5 ignition: ⑴ ignition sources: gas (or other civil combustible gas, the same below) ;
⑵ vents Aperture: 0.6 to 0.8 mm.
Ambient temperature: -10 ℃ ~ 50 ℃.
7, relative humidity: ≤ 85%.
8 Total power consumption: not more than 650W.